ZIMMER BIOMET, INC. G7 10 DEG ARCOMXL LINER 32MM I; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467); Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/10/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was confirmed.The part passed inspection of the outer profile with the exception of some deformed areas where the liners are damaged.This damage likely occurred during the procedure.This.Product likely left biomet control conforming.Review of device history records found these units were released to distribution with no deviations or anomalies.There are warnings in the package insert that this type of event can occur and risks are addressed in risk documentation.This report is being submitted late as it has been identified in remediation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent an initial hip procedure on an unknown date with competitor product.Subsequently, the patient was revised due to unknown reasons.During this surgery, the surgeon had difficulty seating a g7 liner inside a g7 cup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10 DEG ARCOMXL LINER 32MM I
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6948111
MDR Text Key90296849
Report Number0001822565-2017-07042
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model NumberN/A
Device Catalogue Number010000775
Device Lot Number3218602
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/10/2015
Initial Date FDA Received10/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age90 YR
Patient Weight54
-
-