ZIMMER BIOMET, INC. G7 HIGH WALL E1 LINER 36MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467); Fitting Problem (2183); Device-Device Incompatibility (2919)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/02/2015
Event Type  malfunction  
Manufacturer Narrative
Zimmer biomet complaint- (b)(4).Complaint sample was evaluated and the reported event was confirmed.Relayed by the product evaluation: "the part passed inspection of the outer profile with the exception of some deformed areas where the liners are damaged.This damage likely occurred during the procedure.Product likely left biomet control conforming." dhr was reviewed and no discrepancies were found.Root cause was unable to be determined and is currently being investigated under within an internal health hazard evaluation.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted late as it has been identified in remediation.
 
Event Description
During the procedure, a g7 e1 hi-wall liner would not seat into the shell.The surgeon used a ceramic liner to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HIGH WALL E1 LINER 36MM E
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6950965
MDR Text Key90325235
Report Number0001825034-2017-08743
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model NumberN/A
Device Catalogue Number010000935
Device Lot Number3480323
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/03/2015
Initial Date FDA Received10/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-