• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE GPAC FRAME-ASSEMBLED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE GPAC FRAME-ASSEMBLED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P3930A000026
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2017
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the hand pendant needed to be replaced.Per the hill-rom service manual: do a periodic inspection to make sure all bed functions operate correctly, especially the safety features.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the account performs preventative maintenance on their bed.The technician replaced the hand pendant to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head section would self run down.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GPAC FRAME-ASSEMBLED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key6951051
MDR Text Key89933370
Report Number1824206-2017-00427
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Patient
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3930A000026
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-