• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 10-DEGREE LINER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 10-DEGREE LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch 3500a will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that foam was attached to top of packaging, which when opened, kicked the liner in its package to the ground.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed that the inner foam is adhered to the tyvek lid as reported thus confirming the reported event.Evidence of adhesive transfer is present around the full length and width of the cavity flange.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being filled to relay additional information, which was unknown a the time of the initial medwatch.Udi # - (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10-DEGREE LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6951630
MDR Text Key90341431
Report Number0001825034-2017-08468
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/21/2020
Device Model NumberN/A
Device Catalogue Number010000903
Device Lot Number3652575
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2017
Initial Date FDA Received10/16/2017
Supplement Dates Manufacturer Received12/03/2017
10/22/2018
Supplement Dates FDA Received12/22/2017
11/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-