Product event summary: the sheath, 4fc12 with lot 48856, and the data files for the date of the reported event were returned and analyzed.Visual inspection of the sheath showed the device was not intact and had a kink on two points on the shaft.In conclusion, the reported sheath issue has been confirmed through testing.The sheath, 4fc12 with lot 48856, failed the returned product inspection due to kinks on the shaft.
|
It was reported that during a cryo ablation procedure, after inflation; there was deflation and this occurred multiple times.The balloon catheter was replaced.Additionally, here was no possibility to introduce the ¿shift¿ to the vein.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event on (b)(6) 2017 the sheath subsequently tested out of specification per the manufacturer¿s investigation.
|