• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Electrical /Electronic Property Problem (1198); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 09/18/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that fuses were replaced and replaced power conversion due to lack of 96v power for motion.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect power conversion has been received by manufacturer for evaluation but results are not yet available.The suspect fuses have not been received by manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure the image acquisition system (ias) of the imaging system would not turn when pressing on switch but mobile view station (mvs) turned on.Representative mentioned that battery indicator lights were full.Disconnecting the umbilical cable did not resolve the issue.There was no patient present at the time of issue.
 
Manufacturer Narrative
The enclosure power conversion was returned to the manufacturer for analysis.Analysis found that the power conversion assembly failed its functional test.The transistors were found to have shorted.There was an electrical issue identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
pa chia vue
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6952679
MDR Text Key90397455
Report Number1723170-2017-04147
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/18/2017
Initial Date FDA Received10/16/2017
Supplement Dates Manufacturer Received05/03/2018
Supplement Dates FDA Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-