• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 40MM G; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 40MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was not confirmed.The product was examined and found to have scratches, a gouge and some deformation likely caused by removal of the liner during the procedure.The product was checked to print and all dimensions are in tolerance except for the profile which is likely out of tolerance due to the deformation.Review of the mhr determined that as the product likely left zimmer biomet conforming to print, the complaint cannot be confirmed.The root cause cannot be determined.Dhr was reviewed and no discrepancies were found.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during a right total hip arthroplasty, the polyethylene liner would not fully seat in the acetabular shell another liner was used to complete the procedure.No adverse events have been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 40MM G
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6954098
MDR Text Key90535381
Report Number0001825034-2017-08871
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberN/A
Device Catalogue Number010000748
Device Lot Number3508158
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/23/2015
Initial Date FDA Received10/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age71 YR
-
-