• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES OBERDORF OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.021
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device coupling tool side did not function.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that the dial on the oscillating saw attachment device came off.According to the report, the cutter device became loose during use.It was reported in the service order that the device had an unspecified malfunction.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILL-SAW-ATTACHM F/532.001 532.010 532
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6961731
MDR Text Key90556529
Report Number8030965-2017-15619
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K971544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2017
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-