MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2742-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/06/2017
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not available for manufacturer evaluation.As such, a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.A records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There were no non-conformance reports, rework, or abnormalities identified during the manufacturing process which could be associated with the reported event.In addition, the dhr review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.
 
Event Description
A certified clinical hemodialysis technician (ccht) reported that the arterial bloodline tubing was leaking from the connector.This occurred at the end of a patient treatment.The patient was able to complete treatment on the machine with no harm or injury.The amount of blood loss was not provided.The bloodlines was not available to be returned to the manufacturer for physical evaluation.No further information has been made available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6962884
MDR Text Key90663292
Report Number8030665-2017-00862
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100309
UDI-Public00840861100309
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Catalogue Number03-2742-9
Device Lot Number17DR01156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age MO
Date Manufacturer Received09/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-