It was reported that stent migration occurred.The target lesion was located in the left anterior descending (lad) artery.After manually aspirating a visible thrombus in the lad, a 3.00 x 24mm synergy ii drug eluting stent was deployed at 12 atmospheres.The balloon was fully deflated; however, upon attempting to remove the stent delivery system (sds), resistance was encountered and upon pulling sds again, it was noted that the stent moved proximally into the lad and left main segment.The stent was then post-dilated with a balloon catheter and was now well apposed to the lesion.The procedure was completed.No patient complications were reported and the patient's status was fine.
|