The customer stated that they received erroneous results for one patient sample tested for gluc3 glucose hk gen.3 (glu) on two cobas integra 400 plus (i400+) analyzers.An erroneous result was reported outside of the laboratory.The sample was measured several times on two i400+ analyzers and the results were approximately 6 to 7 mg/dl.The sample was measured with a glucometer and the results were approximately 97 to 107 mg/dl.The customer collected a new sample from the patient and repeated measurements on an i400+ analyzer.The customer measured different tubes of this sample consisting of different sample material (serum, plasma, serum collected from capillary blood).All results were comparable.It was asked, but it is not known if any adverse events occurred.No adverse events were alleged to have occurred with the patient.The serial numbers of the i400+ analyzers were asked for, but not provided.
|
The patient sample was sent to another laboratory for testing with a different method, glucose dehydrogenase.The result was comparable to results obtained on the i400+ analyzers.The customer consulted the physician and determined that the low glucose values were probably due to high vitamin c supplementation of the patient or incorrect use of the glucometer.The customer noticed a tendency toward lower blood glucose levels of the patient over a period of time, but there were no symptoms of hypoglycemia.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.No further events have been reported by the customer since the event.No explicit issues could be found with the reagent or analyzer.A customer handling issue may have caused the event.
|