Model Number MMT-523RNAS
Device Problems Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Patient Problem Hyperglycemia (1905)
Event Date 07/19/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The insulin pump had minor scratched lcd window, cracked case near display window corners, cracked battery tube threads, broken reservoir tube lip, reservoir tube cracked, cracked reservoir tube window, missing reservoir tube o-ring and cracked lcd window.The pump was received with broken off reservoir tube lip.Reservoir unable to lock in place due to reservoir tube lip completely broken off.
 
Event Description
The customer reported via phone call that the insulin pump was damaged.The top of the cylinder appeared worn but it was not broken, just the ring came out.Customer's blood glucose level was 300 mg/dl.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6972161
MDR Text Key90457025
Report Number2032227-2017-60241
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169513099
UDI-Public(01)00643169513099
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 10/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523RNAS
Device Catalogue NumberMMT-523RNAS
Device Lot NumberA1523RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/13/2017
Initial Date FDA Received10/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age70 YR
Patient Weight130
-
-