AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-PFO-025
Device Problem Material Deformation (2976)
Patient Problem Tissue Damage (2104)
Event Date 06/28/2017
Event Type  Injury  
Event Description
On (b)(6) 2016, a 25 mm amplatzer pfo occluder (pfo) was implanted.On an unknown date, the pfo waist was found to have deformed and protruded through the device mesh damaging the cardiac tissue.On (b)(6) 2017, the pfo was explanted.
 
Manufacturer Narrative
An event of a deformed 25 mm amplatzer pfo occluder (pfo) was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6976363
MDR Text Key90216881
Report Number2135147-2017-00138
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Model Number9-PFO-025
Device Catalogue Number9-PFO-025
Device Lot Number5239484
Other Device ID Number00811806011288
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2015
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-