ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number LXT-CHINA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/28/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported the liquid entered the gas line during a procedure.There was no harm to the patient.Additional information was requested; however, none has been received to date.This one of three events for this facility.
 
Manufacturer Narrative
This is the eighth complaint for the finish goods lot; however, the third for this reported issue.The device history record review shows the order was built and released per specification.The customer did not retain a sample for this complaint report; visual inspection or functional testing could not be conducted.If a sample is returned at a later date, the investigation will be reopened and the sample evaluated.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.After a thorough investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6978790
MDR Text Key91075371
Report Number2028159-2017-04144
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLXT-CHINA
Device Catalogue Number8065751601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2017
Initial Date FDA Received10/25/2017
Supplement Dates Manufacturer Received12/08/2017
Supplement Dates FDA Received01/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-