• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problems Device Contamination with Body Fluid (2317); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
A complete lead was returned and dried blood was noted throughout the inner coil lumen.Visual and x-ray examination reveal damages that are consistent with those occurring at the time of the implant and explant procedure.Electrical testing did not reveal any indication of conductor fractures or internal shorts.The results of the investigation concluded the analysis of the device was normal with no anomalies found.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During follow-up, the left ventricular lead was found to be dislodged.The lead was revised and during the replacement procedure, blood was noted on the lv channel of the header.The lead was explanted and replaced and the patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6980134
MDR Text Key90335539
Report Number2017865-2017-32986
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model Number1258T/86
Device Lot NumberA000028628
Other Device ID Number05414734502610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-