AV-TEMECULA-CT RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Catalog Number 1011344-40
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 08/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device.The reported cracked luer was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to operational context.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that during preparation, the flush port of a.014 acculink ii self-expanding stent was noted to be cracked when a syringe was attached.The device could therefore not be flushed and it was not used in the anatomy.It was replaced with another acculink device to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.Returned device analysis revealed that there was a leak coming from the cracked flush port of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Type of Device
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6982227
MDR Text Key91074795
Report Number2024168-2017-08501
Device Sequence Number1
Product Code NIM
UDI-Device Identifier08717648076398
UDI-Public08717648076398
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number1011344-40
Device Lot Number7031561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2017
Initial Date FDA Received10/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-