Model Number GIF-Q260J
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to omsc for evaluation.During the evaluation, omsc found no irregularity inside of the instrument channel and suction channel of this device.Although omsc found no irregularity also around the instrument channel port and the cylinder, it was confirmed damage at around the instrument channel outlet of the distal end and peeling of the coat of the insertion tube.Also, omsc reviewed the manufacturing history of this device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of microbiological testing by the user facility, the microbes were detected from the subject device.The type and number of microbes are unknown.There was no report of infection associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key6986295
MDR Text Key91931436
Report Number8010047-2017-01582
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-Q260J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2017
Date Manufacturer Received10/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-