The device was received for evaluation and an evaluation is complete.A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot.A visual inspection was performed with the naked eye.Functional testing was performed which included leak testing, clear passage testing, and clamp function testing.Cut tubing on the supply line, fill line, patient line, drain line, and several cuts to the cassette sheeting were identified during visual inspection.The unit leaked through the cuts on the tubing lines and cassette sheeting.The cause of event was determined to be flow wrap damage during manufacturing.Should additional relevant information become available, a supplemental report will be submitted.
|