Model Number CD3365-40Q
Device Problem No Device Output (1435)
Patient Problem Ventricular Tachycardia (2132)
Event Date 10/06/2017
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital with slow ventricular rate; failure to deliver ventricular tachycardia pacing was observed on the implantable cardioverter defibrillator.Non-sustained right ventricular oversensing episodes were noted, which showed similar behavior where pseudo pseudofusion was observed.Programming modifications were performed successfully.The patient¿s condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6992148
MDR Text Key90765064
Report Number2017865-2017-33928
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2016
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot Number4581697
Other Device ID Number05414734508292
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-