Catalog Number 3MAXC |
Device Problems
Fracture (1260); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the penumbra system 3max reperfusion catheter (3max) was fractured approximately 29.5 cm from the hub.The 3max catheter shaft was twisted.Conclusions: evaluation of the returned device confirmed that the 3max was fractured.This type of damage typically occurs due to improper handling during removal from the packaging.If the device is forcefully retracted an extreme angles during removal from its packaging hoop, damage such as this may occur.Further evaluation revealed that the distal shaft of the 3max was twisted.This damage was likely incidental and likely occurred while handling the device after it was removed from its packaging.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the hospital staff opened and handed off the penumbra system 3max reperfusion catheter (3max) to the physician.Upon being handed the 3max, the physician noticed that the 3max was fractured about thirty centimeters from its the proximal end.The damaged 3max was found prior to use and therefore, was not used in the procedure.The procedure was successfully completed using another microcatheter.
|
|
Manufacturer Narrative
|
Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #01 mfr report: 1.Usage of device.
|
|
Search Alerts/Recalls
|