Catalog Number 544965
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation at this time.Teleflex will continue to monitor and trend related events.
 
Event Description
Reported issue: the applier was not releasing properly during a procedure.Additional information indicates that the applier failed to release the clip.There was no patient injury.
 
Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was manufactured at the (b)(4) as part of a 50pc.Lot in march of 2016.The returned instrument was evaluated and found that the knob rotation and handle to jaw mechanisms were both dry and sluggish feeling signifying that this instrument is in need of proper lubrication.Further evaluation showed that when attempting to close a clip to verify function, the jaw mechanism bound up after clip was released thus we are able to validate the alleged complaint.Parts were 100% visually inspected and tested at the (b)(4) before instruments were sent to the customer.No irregularities were found and or reported at the time of inspection and assembly of the product as this is a standardized process for all instruments manufactured at this facility.At this time it is suspected that this instrument was not lubricated prior to sterilization at the customer's facility as instructed in ifu l6109 r.04 which was supplied with the instrument at the time of manufacture which states "appropriate care, cleaning, lubrication and maintenance are important to ensure proper function." after initial investigation this instrument was properly other remarks: lubricated as instructed in ifu 6109 r.04 and proper smooth function to all of its mechanisms was restored.No corrective action required at this time.
 
Event Description
Reported issue: the applier was not releasing properly during a procedure.Additional information indicates that the applier failed to release the clip.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5 MILLIMETER ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key6997119
MDR Text Key91337468
Report Number3011137372-2017-00349
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2017
Initial Date FDA Received11/02/2017
Supplement Dates Manufacturer Received12/18/2017
Supplement Dates FDA Received12/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-