A review of the manufacturing records for the device verified the lot met all pre-release specifications.The engineering evaluation stated the following: the distal shaft, upon which the endoprosthesis was mounted, and the remainder of the delivery catheter assembly were returned.No endoprosthesis, deployment knob, nor deployment line were returned.The catheter assembly was returned in 3 separate pieces: the distal tip, the distal shaft, and the remainder of the delivery catheter assembly.The distal shaft was approximately 11.5 cm long.Neither end of the distal shaft showed melting or disruption of the pebax coating.Manufacturing anomalies were identified that potentially could have contributed to the event.
|