• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY RAYNHAM ¿ 1219655 SIGMA HP UNI TIB TRAY SZ3 LMRL; KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY RAYNHAM ¿ 1219655 SIGMA HP UNI TIB TRAY SZ3 LMRL; KNEE TIBIAL TRAY Back to Search Results
Catalog Number 102451300
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Pain (1994); Limited Mobility Of The Implanted Joint (2671)
Event Date 10/04/2017
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised to address tibial subsidence, pain and laxity.It was also reported that the patient had also developed some patella femoral arthritis.
 
Manufacturer Narrative
No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB TRAY SZ3 LMRL
Type of Device
KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY RAYNHAM ¿ 1219655
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY RAYNHAM ¿ 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6998239
MDR Text Key90929787
Report Number1818910-2017-27968
Device Sequence Number1
Product Code HRY
UDI-Device Identifier10603295002338
UDI-Public10603295002338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number102451300
Device Lot Number641622
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2017
Initial Date FDA Received11/02/2017
Supplement Dates Manufacturer Received11/20/2017
Supplement Dates FDA Received11/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient Weight128
-
-