SYNTHES OBERDORF SAGITT SAW-ATTACHM LONG W/T-HANDLE F/TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.224 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was determined that the coupling power supply was loose on the sagittal saw attachment device.It was further determined that the coupling pins came apart.It was further determined that the device failed pretest for check pins length of cone sleeve.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|