(b)(4).Pma/510(k) #p100022/s001.This investigation captures the restenosis of a ziv6-35-125-7.0-120-ptx (lot# c777750) stent placed in the left sfa of a male patient on (b)(6) 2012.On (b)(6) 2017 restenosis in the stented lesion was confirmed.Worsened intermittent claudication was observed on the patient.Pta was conducted against the restenosis on the same day and the condition of the patient recovered.The ziv6-35-125-7.0-120-ptx stent of lot number c777750 was implanted in the patient and is therefore unavailable for evaluation.With the information provided a document based investigation was carried out.It was stated by the originator that there are no images available for review.It is known the patient had the following pre-existing condition: coronary artery disease there is no evidence to suggest that this event did not occur, therefore the complaint is confirmed based on customer testimony.Restenosis is a common adverse event of endovascular procedures and can be caused by injury to the vessel (e.G.During percutaneous transluminal angioplasty (pta) and/or stenting).Vessel injury provokes an inflammatory response that leads to (or amplifies) the restenosis process.It may be noted that surface of the zilver ptx stent is coated with the drug (paclitaxel) to help prevent subsequent restenosis of the artery.It can be therefore stated that it is very unlikely that the reported restenosis could have occurred due to zilver ptx malfunction; however as no imaging was available for review, a definitive root cause of this event cannot be determined.It may be noted that the packaging insert lists restenosis of the stented artery as a known potential adverse event associated with placement of this device.In addition worsened claudication is also listed as a potential adverse event associated with the placement in this device as per the packaging insert.Prior to distribution all zilver ptx devices are subject to visual inspection and functional checks to ensure device integrity.A review of the relevant manufacturing records revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has occurred previously with the lot number c777750.However, based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with the lot number c777750.Therefore no action required.Pta was conducted against the restenosis on the same day and the condition of the patient recovered.Complaints of this nature will continue to be monitored for any potential emerging trends.
|