According to the customer: "15 min after the start of the heart-lung machine sudden delta p rise above 200mmhg.Measured pressure before oxy 400mmhg, measured pressure according to oxy 180mmhg.Pao2 drop from 240mmhg to 130mmhg.Act during the event always over 430 acute change of the oxygenator (emergency situation), aterial mean pressure of the patient during the oxygenator change to 30 mmhg, no long-term sequences visible to the patient." ref.: #(b)(4).
|
The product was visually inspected in the laboratory of the manufacturer.On the blood inlet and blood outlet side no clots were detected.During rinsing no clots were flushed out.The product was cleaned with sodium hypochlorite solution.No other abnormalities were detected.Dhr review: affected product: basic lot 70115732 and packaging lot 70115696 (serial number (b)(4).The avz from gkv 820 to gkv 834 (b)(4) was reviewed on (b)(6) 2017.There were no references found, which are indicating a nonconformance of the product in question.Additionally a review of the weekly performance monitoring according to si-c-09 has been reviewed (b)(4), the performance tests passed the acceptance criteria.The reported failure could not be confirmed.Based on the results obtained during investigation at this time the cause of the reported incident was determined to not be attributed to a device related malfunction since the reported failure did not contribute to a death or serious injury and no systemic issue could be determined no corrective action is needed at this time.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
|