• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER Back to Search Results
Catalog Number GSX0030A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Obstruction/Occlusion (2422)
Event Date 10/12/2017
Event Type  Injury  
Manufacturer Narrative
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device was discarded, so no engineering investigation could be performed.(b)(4).
 
Event Description
It was reported the physician was implanting a 30 mm gore® cardioform septal occluder in an atrial septal defect that balloon sized to 16x17 mm.The device was deployed with good apposition to the septum.Following the procedure the pediatric patient woke from anesthesia in an unrelaxed state.Imaging showed the device embolized to the main pulmonary artery.The patient was sent to surgery for removal of the device and closure of the defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
TRANSCATHETER, SEPTAL OCCLUDER
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
ashley marostica
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key7004312
MDR Text Key91141177
Report Number2017233-2017-00580
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/05/2019
Device Catalogue NumberGSX0030A
Device Lot Number16171214
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age6 YR
Patient Weight15
-
-