Model Number 3CXSX25R |
Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/20/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.(b)(4).Results code: results pending completion of evaluation.Conclusions code: conclusion not yet available-evaluation in progress.
|
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the water line arm was cracking and breaking off.No patient involvement.Product was changed out.Procedure was completed successfully.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.The returned sample was visually inspected.It was found that 5 of the 6 heat exchanged water port l-tubes were broken off.Additionally, scuff marks were noted on each returned he housing.A retention sample from the same product code/lot number combination was visually inspected and no anomalies were noted.It is likely that the units were exposed to damage after manufacturing that caused the water tubes to break.It was not able to be determined how or when the damage occurred.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|