MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SX OXYGENATOR W/HR (STERILE); BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SX OXYGENATOR W/HR (STERILE); BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CXSX25R
Device Problems Break (1069); Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 10/20/2017
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.(b)(4).Results code: results pending completion of evaluation.Conclusions code: conclusion not yet available-evaluation in progress.
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the water line arm was cracking and breaking off.No patient involvement.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.The returned sample was visually inspected.It was found that 5 of the 6 heat exchanged water port l-tubes were broken off.Additionally, scuff marks were noted on each returned he housing.A retention sample from the same product code/lot number combination was visually inspected and no anomalies were noted.It is likely that the units were exposed to damage after manufacturing that caused the water tubes to break.It was not able to be determined how or when the damage occurred.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SX OXYGENATOR W/HR (STERILE)
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key7006233
MDR Text Key92200714
Report Number1124841-2017-00232
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model Number3CXSX25R
Device Catalogue NumberN/A
Device Lot NumberUM03
Other Device ID Number(01)00699753450073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/20/2017
Initial Date FDA Received11/07/2017
Supplement Dates Manufacturer Received12/19/2017
Supplement Dates FDA Received12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-