ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER; GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY Back to Search Results
Catalog Number ULT10.2-38-40-P-32S-CLB-RH
Device Problem Leak/Splash (1354)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/13/2015
Event Type  malfunction  
Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event Description
After the drain was placed, it began to leak at the junction between the catheter and the hub.It was needed to be screwed back together.A section of the device remained inside the patients body was it was intended as the initial drainage catheter was removed and replaced with a new catheter successfully.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER
Type of Device
GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key7007757
MDR Text Key92591427
Report Number1820334-2017-03940
Device Sequence Number1
Product Code GCA
UDI-Device Identifier00827002094987
UDI-Public(01)00827002094987(17)171104(10)5415317
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberULT10.2-38-40-P-32S-CLB-RH
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/07/2017
Initial Date FDA Received11/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-