It was reported that during placement of the 4th embolization coil, resistance was encountered.During removal, the coil unexpectedly detached.Both segments of the coil were removed from the patient.There was no reported intervention or patient injury.The current patient's status is reported to be stable.
|
The pusher and introducer were returned for analysis.The body coil was noted to be stretched, starting from platinum coil.The implant was stretched, with the monofilament broken at the tie knot portion.No damage was noted on the remainder of the pusher, and appeared to be in conformance; however, the second half of the implant was not returned for analysis.Based upon the investigation findings and available information, the complaint of a broken coil was confirmed.The root cause could not be determined; however, the device exhibited evidence that it was subjected to forces that exceeded its strength specification, evident by the stretched coil and broken monofilament.
|