• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER Back to Search Results
Model Number 16402
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This unit is a test unit and not used clinically.The fsr installed a new dual tach printed circuit board (pcb).The unit operated to the manufacturer's specifications.The defective part was returned to the manufacturer for further evaluation.This complaint is related to (b)(4) / medwatch #1828100-2017-00509.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the roller pump displayed a "belt slip" error message.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) installed the dual tach board into an 8k pump test fixture.The pst powered on the pump and started rotations of the pump.No belt slip errors occurred.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7010101
MDR Text Key92617781
Report Number1828100-2017-00510
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-