Model Number 16402 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This unit is a test unit and not used clinically.The fsr installed a new dual tach printed circuit board (pcb).The unit operated to the manufacturer's specifications.The defective part was returned to the manufacturer for further evaluation.This complaint is related to (b)(4) / medwatch #1828100-2017-00509.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the roller pump displayed a "belt slip" error message.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) installed the dual tach board into an 8k pump test fixture.The pst powered on the pump and started rotations of the pump.No belt slip errors occurred.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|