ANATOMAGE ANATOMAGE GUIDE; SURGICAL GUIDE Back to Search Results
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Perforation (2001)
Event Date 10/12/2017
Event Type  Injury  
Manufacturer Narrative
Suspect device was not returned.Process evaluation did not reveal any production error.Review of the treatment plan showed that the implant was planned lingually and centered in the bone.Free handing the surgery to make it more buccal caused the implant to perforate the buccal bone plate.
 
Event Description
Doctor used a surgical guide to place dental implant.After the pilot drill, doctor thought that the trajectory was very lingual in comparison to how he planned the treatment plan.Doctor freehanded the surgery but ended up perforating the buccal plate instead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOMAGE GUIDE
Type of Device
SURGICAL GUIDE
Manufacturer (Section D)
ANATOMAGE
303 almaden blvd
suite 700
san jose CA 95110
Manufacturer (Section G)
ANATOMAGE
303 almaden blvd
suite 700
san jose CA 95110
Manufacturer Contact
phyo soe
303 almaden blvd
suite 700
san jose, CA 95110
4088851474
MDR Report Key7010400
MDR Text Key91360802
Report Number3008272529-2017-00033
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 10/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age75 YR
-
-