• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 40X1; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 40X1; BONE CEMENT, ANTIBIOTIC Back to Search Results
Model Number N/A
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign.The event occurred in (b)(4).The product within this report is a combination product.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported the cement powder pouch was not sealed properly, allowing the cement powder to spill out.A delay of 10 minutes was reported.No additional patient consequence or further information have been reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Evaluation of returned device found the supplier sealing was opened partial length.The manufacture sealing was intact.Device history record (dhr) was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to supplier packaging issue.Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN BONE CEMENT R 40X1
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7010527
MDR Text Key92211860
Report Number3006946279-2017-00236
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
PK150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/31/2019
Device Model NumberN/A
Device Catalogue Number4003940001
Device Lot NumberB608AA1904
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-