Product event summary: the sheath, 4fc12 with lot 24668, was returned and analyzed.Visual inspection of the sheath showed the device shaft was kinked 0.9270 inches from the tip.In conclusion, the reported issue was confirmed through testing.The sheath failed the returned product analysis due to a shaft kink near the tip.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the steering mechanism on the sheath was broken.The sheath was not replaced.The case was completed with cryo.No patient complications have been reported as a result of this event.(b)(6) 2017 the sheath subsequently, tested out of specification per the manufacturer's investigation.
|