Model Number 1458Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problems Syncope (1610); Fall (1848)
Event Date 10/19/2017
Event Type  Injury  
Event Description
It was reported that the patient was seen in hospital after passing out during a ventricular tachycardia episode; the device appropriately treated the arrhythmia.The left ventricular lead exhibited loss of capture; x-ray revealed the lead was dislodged; the physician suspected that this was due to the patient suffering a fall during the vt episode.The lead was explanted and replaced on (b)(6) 2017.The patient was in stable condition post procedure.
 
Manufacturer Narrative
Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7024992
MDR Text Key91807125
Report Number2017865-2017-34492
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model Number1458Q/86
Device Lot NumberA000044184
Other Device ID Number05414734503198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-