Catalog Number ULT12.0-38-40-P-32S-CLB-RH
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/29/2015
Event Type  malfunction  
Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol_cap008, compliant/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event Description
The patient had a 12fr biliary drainage tube placed under general anesthesia.The tube was flushed and injected with contrast and no leak was seen.The patient was discharged.Approximately 3 hours later the patient called stating the tube was leaking.The facility requested the patient come in to the imaging department.The tube was examined by the lead tech and the physician.There was an obvious leak at the junction of the drainage tube and mac-loc.The physician was able to unscrew the tube from the mac-loc and thread it back on properly.The retention filament was cut to accomplish this.The tube did not leak after that.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER
Type of Device
GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key7025617
MDR Text Key92765361
Report Number1820334-2017-04019
Device Sequence Number1
Product Code GCA
UDI-Device Identifier00827002094994
UDI-Public(01)00827002094994(17)180609(10)5945594
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberULT12.0-38-40-P-32S-CLB-RH
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/07/2017
Initial Date FDA Received11/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-