Model Number 20027E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/23/2017
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: non-bd spike adaptors; (2)spiro connectors, therapy date: (b)(6) 2017.Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
 
Event Description
The customer reported that the filter leaked during patient use.There was no report of patient harm.
 
Manufacturer Narrative
(b)(4).The customer¿s report of a leaking filter was confirmed.Visual inspection of the set did not reveal any discernible damage or abnormalities.Microscopic examination of the leaking component did not reveal any noticeable damage to either the filter membrane or housing.Functional testing confirmed leaking from the upper vent hole of the filter.The root cause of the leaking was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS EXTENSION SET
Type of Device
SET, EXTENSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key7026245
MDR Text Key92546458
Report Number9616066-2017-01517
Device Sequence Number1
Product Code FPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K801614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20027E
Device Catalogue Number20027E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-