MEDTRONIC SOFAMOR DANEK USA, INC PIVOX OBLIQUE LATERAL SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 2140010
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 10/17/2017
Event Type  Injury  
Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event Description
It was reported that patient underwent an oblique lumbar inter-body fusion surgery (olif) at l3-l4.Intra-op, inserter's screw at the tip was broken.The broken piece was stuck inside the locking cap of the cage.Surgeon was unable to twist the locking cap with the screw driver as the broken piece was stuck in the middle of the locking cap.No patient complications were reported as a result of the event.
 
Manufacturer Narrative
X-ray review images for l3-4, l4-5, l5-s anterior lumbar fusion was provided.On the magnified image at l3-4, there is an opacity in the locking cap presumably a broken piece of the plate inserter.It is unclear if this is related to technique or a defect in the inserter release mechanism.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVOX OBLIQUE LATERAL SPINAL SYSTEM
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key7026506
MDR Text Key91877687
Report Number1030489-2017-02321
Device Sequence Number1
Product Code MAX
UDI-Device Identifier00643169639522
UDI-Public00643169639522
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K152277
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2140010
Device Lot NumberNM16E086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-