Catalog Number 26-1222
Device Problem Failure to Auto Stop (2938)
Patient Problem Tissue Damage (2104)
Event Date 10/22/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
As reported by the ous affiliate, 2 perforators did not disengage, resulting in damage to the dura.
 
Manufacturer Narrative
Device was returned for evaluation.A follow up report will be filed upon completion of the investigation.
 
Manufacturer Narrative
(b)(4).The device was returned for evaluation.The perforator was visually inspected.No anomalies were found.The device was functionally tested.No issues were noted.The device functioned as intended.A review of manufacturing records found no anomalies during the manufacturing process.Based on the evaluation of the device, the reported issue could not be confirmed.The unit functioned as intended.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.Refer to mdr 1226348-2017-10829 for information regarding the second device associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISP PERFORATOR 11MM
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key7026648
MDR Text Key91902847
Report Number1226348-2017-10830
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number26-1222
Device Lot NumberHB1457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/23/2017
Initial Date FDA Received11/13/2017
Supplement Dates Manufacturer Received11/20/2017
03/12/2018
Supplement Dates FDA Received11/20/2017
03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-