Catalog Number LSM0800826 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/26/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of device is not expected.The investigation is currently in progress.
|
|
Event Description
|
It was reported that the stent moved when the device was inflated.The device was used as off label use for a fenestration in a cook aaa stent graft procedure.The stent was positioned as intended, the second strut segments inside the fenestration and the rest out of the fenestration.When the lifestream balloon was inflated the stent moved distally and the destination of the expanded stent was different from the position before inflation.A compliant flaring balloon was then used to try to flare the stent.After that when the control angiogram was performed it showed that the full stent body had landed outside the aaa graft window.Therefore it did not provide a sealed blood flow to the visceral artery as required.To fix this a 7mm atrium v12 stent was used and inflated proximal to the lifestream.It was then successfully flared and could provide the sealed blood flow to the visceral artery.There was no patient injury reported.
|
|
Manufacturer Narrative
|
It was reported that the stent moved when the device was inflated.There were no anomalies noted during the prep of the device.There was no damage noted to the box, tray or stent cover and there was no difficulty removing the stent cover.A cook sheath and a 0.035 guidewire were used.Femoral access was made.The target lesion was the left renal artery.The device was used as off label use for a fenestration in a cook aaa stent graft procedure.There was an old aortic stent graft in place at the target lesion and the cook fever was placed over it.The stent was positioned as intended, the second strut segments inside the fenestration and the rest out of the fenestration.When the lifestream balloon was fully inflated the stent moved distally and the destination of the expanded stent was different from the position before inflation.A compliant flaring balloon was then used to try to flare the stent.After that when the control angiogram was performed it showed that the full stent body had landed outside the aaa graft window.Therefore it did not provide a sealed blood flow to the visceral artery as required.Optimal stent location was performed prior to inflating the device.To fix this a 7mm atrium v12 stent was used and inflated proximal to the lifestream.It was then successfully flared and could provide the sealed blood flow to the visceral artery.Guidewire access was maintained throughout the procedure.There was no patient injury reported.The lot history records have been reviewed with special attention to the manufacturing and inspection of this product.The device was found to have met specifications prior to shipment.No manufacturing anomalies were identified that may have caused or contributed to the reported event.This is the first reported complaint for this lot number and issue to date.The device was not returned for evaluation.The result of the investigation is inconclusive.The sample was not returned for evaluation.The event described that the lifestream device was used in an off label fenestration in a cook aaa stent graft.The lifestream stent is intended only in the treatment of atherosclerotic lesions in the common and external iliac arteries as per lifestream ifu.Based upon the available information a definitive root cause has not been determined.Based on analysis performed no additional action is required at this time.The ifu states: device description: implant the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from 316l stainless steel, encapsulated between two layers of eptfe.Indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation: using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection: select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation: carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation a 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent advance the endovascular system over the guidewire into the introducer sheath.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.(b)(4).
|
|
Event Description
|
It was reported that the stent moved when the device was inflated.There were no anomalies noted during the prep of the device.There was no damage noted to the box, tray or stent cover and there was no difficulty removing the stent cover.A cook sheath and a 0.035 guidewire were used.Femoral access was made.The target lesion was the left renal artery.The device was used as off label use for a fenestration in a cook aaa stent graft procedure.There was an old aortic stent graft in place at the target lesion and the cook fever was placed over it.The stent was positioned as intended, the second strut segments inside the fenestration and the rest out of the fenestration.When the lifestream balloon was fully inflated the stent moved distally and the destination of the expanded stent was different from the position before inflation.A compliant flaring balloon was then used to try to flare the stent.After that when the control angiogram was performed it showed that the full stent body had landed outside the aaa graft window.Therefore it did not provide a sealed blood flow to the visceral artery as required.Optimal stent location was performed prior to inflating the device.To fix this a 7mm atrium v12 stent was used and inflated proximal to the lifestream.It was then successfully flared and could provide the sealed blood flow to the visceral artery.Guidewire access was maintained throughout the procedure.There was no patient injury reported.
|
|