Results: the penumbra system 026 separator flex (separator 026 flex) delivery wire was fractured approximately 42.0 cm from the proximal end, with slight bending of the wire just proximal and distal to the fracture.The segment distal to the fracture was kinked approximately 1.0 cm distal to the fracture.There was no visible damage to the separator bulb or atraumatic tip.Conclusions: evaluation of the returned device revealed the delivery wire was fractured and kinked.This damage may have occurred due to forceful handling of the separator 026 flex at extreme angles during advancement through a catheter.There was no visible damage to the separator bulb or atraumatic distal tip.The catheter 026 mentioned in the complaint was not returned for evaluation.Penumbra separators are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
While preparing a penumbra system 026 separator flex (separator 026 flex) on the back table for a thrombectomy procedure, the tip of the separator 026 flex broke off.It was reported that the separator 026 flex was being prepped through a penumbra system reperfusion catheter 026 (catheter 026) and that nothing out of the ordinary was noticed.The tip of the separator 026 flex broke prior to use and therefore, the separator 026 flex was not used in the procedure.The procedure was successfully completed using a new separator 026 flex and the same catheter 026.
|