Model Number H74939032504080 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/31/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).
|
|
Event Description
|
It was reported that air was present in the balloon.A 5.0mmx40mmx80cm (4f) sterling¿ balloon catheter was selected for use to dilate the lesion.However, during preparation, when air removal was performed, there was an air present on the proximal side of the balloon.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The balloon was loosely folded.There is contrast media in the balloon and in the inflation lumen.The presence of contrast media in the lumen is indicative of handling beyond simply unpackaging the device.The outer shaft, inner shaft, balloon and tip were microscopically examined.Inspection of the device presented no damage or irregularities.Functional testing was carried out by attaching an inflation device filled with water to the hub of the complaint device, and inflating the device to rated burst pressure.The sterling device inflated and held pressure for 5 minutes without leaking.There was no evidence of any material or manufacturing deficiencies contributing to the reported event.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
|
Event Description
|
It was reported that air was present in the balloon.A 5.0mmx40mmx80cm (4f) sterling¿ balloon catheter was selected for use to dilate the lesion.However, during preparation, when air removal was performed, there was an air present on the proximal side of the balloon.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Search Alerts/Recalls
|