Model Number BI-700-00027-120
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913); Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 07/28/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The representative reported that the gantry motion controller was replaced without resolution.It was noted that further troubleshooting found that the winch assembly required replacement.However, confirmation of replacement has not been provided.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the gantry door of the imaging system could not be completely opened.It was reported that the motor could not complete the opening process.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was reported that the returned door winch was lost at the mdt brazil warehouse and is unavailable for further analysis.The manufacturer representative also provided event information stating that before the replacement of the door winch, it was possible to open the gantry door with the manual wrench; meaning that it was possible to completely open the gantry door manually.After replacing the door winch the issue was resolved and the system was completely functional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7039777
MDR Text Key92933139
Report Number1723170-2017-04717
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-