STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK HANDLE, 45CM; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250181164
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/23/2017
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
Gtin: (b)(4).Alleged failure: failed insulscan the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.The product was returned for investigation and the failure mode was confirmed and will be monitored for future reoccurrence.
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM PEEK HANDLE, 45CM
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
ashley lower
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7043041
MDR Text Key93198427
Report Number0002936485-2017-01131
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250181164
Device Lot Number1540431¿H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-