The hill-rom technician performed troubleshooting and found the caster at the head of the bed on the patient's right side needed to be replaced.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the brake casters to determine if the bed moves when you activate the brake, replace or adjust when necessary.Check the tires for cuts, wear, tread life, etc.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The hill-rom technician replaced the caster at the head of the bed on the patient's right side to resolve the issue.Based on this information, no further action is required.
|