Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the screw used to fix the shell would not insert.The surgery was finished with a backup screw.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the screw found minimal damage, scratches were identified on the taper of the head.No damage was observed to the head or threads.Dimensional analysis was performed against the print specifications.The head diameter, major thread diameter, and overall length of the screw were found to be within tolerance.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being filled to relay additional information, which was unknown at the time of the initial medwatch.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 35MM
Type of Device
SCREW, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7044527
MDR Text Key93195216
Report Number0001825034-2017-10201
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number010001000
Device Lot Number6077771
Other Device ID NumberSEE NARATIVE IN H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2017
Initial Date FDA Received11/20/2017
Supplement Dates Manufacturer Received01/04/2018
02/28/2018
Supplement Dates FDA Received01/29/2018
03/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-