Catalog Number BI70000028230 |
Device Problems
Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The representative reported that the generator covers were adjusted and home was invalidated to resolve the reported issue.The imaging system then passed the system checkout and was found to be fully functional.No parts were replaced, therefore no parts were returned for analysis.
|
|
Event Description
|
A site representative reported that, while outside of a procedure, the gantry of the imaging system was unable to complete motions in the x-direction.It was reported that the gantry would move about ten centimeters and then stop.There was no patient present when this issue was identified.No additional information was provided.
|
|
Manufacturer Narrative
|
Correction: unique device identification (udi) updated to proper value.
|
|