• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, ROTATING HANDLE, INSULATED SHAFT; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, ROTATING HANDLE, INSULATED SHAFT; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080231
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/30/2017
Event Type  malfunction  
Manufacturer Narrative
Alleged failure: coating peeling off handle.Confirmed failure: coating peeling off handle.Probable root cause: poor autoclave reliability.Incorrect sterilization/reprocessing procedure.Handling procedures.Contact forces.Product used beyond defined useful life.The device manufacture date is not known.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Gtin: (b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, ROTATING HANDLE, INSULATED SHAFT
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
roomi banerjee dua
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7051425
MDR Text Key93207442
Report Number0002936485-2017-01139
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080231
Device Lot Number¿0842588D¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2017
Date Manufacturer Received10/30/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-