• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR MODULAR HEAD 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR MODULAR HEAD 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74121350
Device Problems Naturally Worn (2988); Appropriate Term/Code Not Available (3191)
Patient Problems Failure of Implant (1924); Pain (1994)
Event Date 05/31/2017
Event Type  Injury  
Event Description
It was reported that right hip revision surgery was performed due to failed right hip arthroplasty secondary to metal-on-metal wear and bony erosion causing pain.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision the modular head was removed.The bhr cup and stem remained implanted.As of today, device return and additional information has been requested for this complaint but has not become available.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.The available medical documents were reviewed.The intraoperative findings of metallosis, osteolysis, and elevated metal ions are consistent with findings associated with metal debris, however the source of the reported clinical reactions cannot be confirmed.It cannot be concluded that the reported reactions/events were associated with a mal-performance of the implant.The patient impact beyond the revision pain/healing cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR MODULAR HEAD 50MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK   CV31 3HL
Manufacturer Contact
markus poettker
schachenallee 29
aarau 5001
SZ   5001
MDR Report Key7051949
MDR Text Key92719325
Report Number3005975929-2017-00402
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061243
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Attorney
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2011
Device Catalogue Number74121350
Device Lot Number50427
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/20/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FEMORAL HEAD, PART AND LOT # UNKNOWN
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
-
-