Model Number 5596
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2017
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The cause of the reported incident could not be conclusively determined, but based on the information received, it was likely due to the use of electrocautery.
 
Event Description
During device replacement for normal eri, the device went into back-up mode due to the use of electrocautery.An external pacing system analyzer (psa) was used during the substitution and the patient experienced symptoms of dizziness at this time.The device was explanted and replaced and the patient was stable following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRONTIER II
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7052965
MDR Text Key92765032
Report Number2017865-2017-34921
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2011
Device Model Number5596
Device Lot Number3102033
Other Device ID Number05414734006941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-